Instrument Make and Model: SBE48 Hull Mounted Temperature Sensor S/N: 480019 Calibration Date: February 19, 2016 installed: March 7, 2016 Installation: installed below the settee, fwd of the galley DD: 20130527 RAW datafile: dsLog time stamp: SSW 2013/01/30 14:00:04.078 SBE48 SSW Science Salt Water 2013/01/30 date in gmt 14:00:04.078 time in gmt SBE48 data collected using SBE48 data 15.2570, 28 May 2013, 00:16:26 26.9851, 30 Jan 2013, 14:00:19 Data Parsed: 26.9851 ttt.tttt temperature (degrees Celsius, ITS-90) 30 Jan 2013 date 14:00:19 time Notes: The instrument sends out two lines of data, the first is the command for 'send data'. The second line is the data. SSW 2013/01/30 14:00:04.078 SBE48 S� SSW 2013/01/30 14:00:04.209 SBE48 # 26.9851, 30 Jan 2013, 14:00:19 Calibration Info from Instrument S>DC SBE48 V 2.0a 00019 temperature: 19-feb-16 TA0 = -2.407645e-04 TA1 = 3.173262e-04 TA2 = -5.178047e-06 TA3 = 2.254826e-07 S> SBE 48 V 2.0a SERIAL NO. 0019 18 Mar 2016 10:19:36 battery voltage = 12.1 logging not started sample interval = 0 seconds number of samples to average = 4 samplenumber = 0, free = 4699867 serial sync mode disabled real-time output enabled SBE 39 configuration = temperature only binary upload does not include time temperature = 5.55 deg C S>startnow S> # 5.5515, 18 Mar 2016, 10:2