Instrument Make and Model: SBE48 Hull Mounted Temperature Sensor S/N: 4824310-0013 Calibration Date: November 17, 2012 installed: May 28, 2013 Installation: installed below the settee, fwd of the galley DD: 20130527 RAW datafile: dsLog time stamp: SSW 2013/01/30 14:00:04.078 SBE48 SSW Science Salt Water 2013/01/30 date in gmt 14:00:04.078 time in gmt SBE48 data collected using SBE48 data 15.2570, 28 May 2013, 00:16:26 26.9851, 30 Jan 2013, 14:00:19 Data Parsed: 26.9851 ttt.tttt temperature (degrees Celsius, ITS-90) 30 Jan 2013 date 14:00:19 time Notes: The instrument sends out two lines of data, the first is the command for 'send data'. The second line is the data. SSW 2013/01/30 14:00:04.078 SBE48 S� SSW 2013/01/30 14:00:04.209 SBE48 # 26.9851, 30 Jan 2013, 14:00:19 The saltwater intake is located in the forward transducer well on the portside at 5 meters below the waterline. Calibration Info from Instrument DS BE48 V 1.0 SERIAL NO. 0013 28 May 2013 00:08:09 not logging: received stop command sample interval = 5 seconds number of samples to average = 12 samplenumber = 0, free = 299593 serial sync mode disabled real-time output enabled temperature = 15.35 deg C S> S> DC SBE48 V 1.0 0013 temperature: 17-nov-12 TA0 = -2.771682e-05 TA1 = 2.801842e-04 TA2 = -2.614672e-06 TA3 = 1.666984e-07 rtc: 01-aug-03 RTCA0 = 0.000000e+00 RTCA1 = 0.000000e+00 RTCA2 = 0.000000e-41 S>