       2005 cruises CTD Fluorescence correction

Processed CTD fluorescence should be recalculated using the formula:

      Fluor_corr = Fluor - (A1 + 1) * A0, where:
      
Fluor is processed CTD fluorescence;
Fluor_corr is corrected CTD fluorescence;
A0 and A1 are calibration coefficients for Fluorometer, Wetlab ECO-AFL/FL - Vblank and Scale factor. The A0, A1 values for the Oceanus OC412 and all Tioga cruises are:

OC412:
A0=0.193, A1=13.426;

Tioga cruises:
A0=0.074, A1=11.