Date: Mon, 2 Aug 2010 12:05:56 -0400 From: Marshall Swartz To: Oceanus Sssg Cc: Marshall Swartz Subject: Re: ctd_484 pressure sensor s/n 69685 Parts/Attachments: 1 OK 19 lines Text 2 Shown ~21 lines Text ---------------------------------------- Laura, With the Paroscientific pressure sensors, our procedure has been to perform two point pressure confirmations at atmospheric and at 6245 dbars to confirm slope and offset. I'm not in the lab to check the date but this was done last year and was within spec and corroborated the 2002 cal. This SBE9 is to be confirmed again in September. Marshall On Mon, Aug 2, 2010 at 10:08 AM, Oceanus Sssg wrote: Hi Marshall.. Dennis was wondering if there was a more recent calibration check for the pressure sensor on ctd 484. The calibration sheet we have is for 1998 and 2002. Thanks, laura