Date: 05/29/2009 Instrument configuration file: C:\data\oc451.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 Surface PAR voltage added : Yes NMEA position data added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 4303 Calibrated on : 2008-09-25 G : 4.38528621e-003 H : 6.48020952e-004 I : 2.22661253e-005 J : 1.71921109e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 2707 Calibrated on : 2008-09-24 G : -1.07240735e+001 H : 1.55862776e+000 I : -1.08232113e-003 J : 1.68742220e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 69685 in 090484 vertical orientation Calibrated on : 12/18/2002 C1 : -4.680530e+004 C2 : -8.320595e-001 C3 : 1.450810e-002 D1 : 3.903100e-002 D2 : 0.000000e+000 T1 : 3.038361e+001 T2 : -5.411459e-004 T3 : 4.085080e-006 T4 : 2.564540e-009 T5 : 0.000000e+000 Slope : 0.99995000 Offset : 0.00000 AD590M : 1.280820e-002 AD590B : -9.210198e+000 4) Frequency 3, Temperature, 2 Serial number : 4312 Calibrated on : 2008-09-23 G : 4.38277854e-003 H : 6.51524980e-004 I : 2.33784325e-005 J : 1.90202508e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 2768 Calibrated on : 2008-09-26 G : -1.06232807e+001 H : 1.51609949e+000 I : -8.62932773e-004 J : 1.48525286e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Fluorometer, Wetlab ECO-AFL/FL Serial number : FLNTURTD-1013 Calibrated on : 2008-04-18 Vblank : 0.0800 Scale factor : 6.00000000e+000 7) A/D voltage 1, User Polynomial Serial number : FLNTURTD-1013 Calibrated on : 2008-04-18 Sensor name : WetLabs Turbidity A0 : -0.14600000 A1 : 2.00000000 A2 : 0.00000000 A3 : 0.00000000 8) A/D voltage 2, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : CST-1117DR Calibrated on : 2008-04-30 M : 21.7010 B : -1.3670 Path length : 0.250 9) A/D voltage 3, Free 10) A/D voltage 4, Altimeter Serial number : 1133 Calibrated on : Scale factor : 14.950 Offset : 0.200 11) A/D voltage 5, Oxygen, SBE 43 Serial number : 0794 Calibrated on : 2009-03-13 Equation : Murphy-Larson Coefficients for Owens-Millard: Soc : 4.3770e-001 Boc : 0.0000 Offset : -0.5792 Tcor : 0.0004 Pcor : 1.35e-004 Tau : 0.0 Coefficients for Murphy-Larson: Soc : 3.93700e-001 Offset : -4.91300e-001 A : -1.70740e-003 B : 1.44320e-004 C : -2.69760e-006 E : 3.60000e-002 Tau : 2.45000e+000 12) A/D voltage 6, User Polynomial, 2 Serial number : Flashbird-5 Calibrated on : Sensor name : Flashbird A0 : 0.00000000 A1 : 1.00000000 A2 : 0.00000000 A3 : 0.00000000 13) A/D voltage 7, Oxygen, SBE 43, 2 Serial number : 0072 Calibrated on : 2009-05-13 Equation : Murphy-Larson Coefficients for Owens-Millard: Soc : 0.0000e+000 Boc : 0.0000 Offset : 0.0000 Tcor : 0.0000 Pcor : 0.00e+000 Tau : 0.0 Coefficients for Murphy-Larson: Soc : 4.30200e-001 Offset : -5.30700e-001 A : -1.49040e-003 B : 1.23170e-004 C : -2.44280e-006 E : 3.60000e-002 Tau : 2.12000e+000 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : Calibrated on : Conversion factor : 0.00000000 Ratio multiplier : 0.00000000