File: 090484_20080312_Conreport-augment.txt Date: 03/12/2008 Instrument configuration file: C:\iFolder\work\Caldata\confiles\090484\090484_20080312.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 Surface PAR voltage added : Yes NMEA position data added : No Scan time added : No 1) Frequency 0, Temperature Serial number : 2265 Calibrated on : 2007-10-24 G : 4.33186392e-003 H : 6.43910172e-004 I : 2.36290352e-005 J : 2.24426517e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 2304 Calibrated on : 2007-10-25 G : -1.04623113e+001 H : 1.58182324e+000 I : -3.45795042e-003 J : 3.62906198e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 69685 in 090484 vertical orientation Calibrated on : 12/18/2002 C1 : -4.680530e+004 C2 : -8.320595e-001 C3 : 1.450810e-002 D1 : 3.903100e-002 D2 : 0.000000e+000 T1 : 3.038361e+001 T2 : -5.411459e-004 T3 : 4.085080e-006 T4 : 2.564540e-009 T5 : 0.000000e+000 Slope : 0.99995000 Offset : 0.00000 AD590M : 1.280820e-002 AD590B : -9.210198e+000 4) Frequency 3, Temperature, 2 Serial number : 2271 Calibrated on : 2007-10-24 G : 4.33366749e-003 H : 6.41047311e-004 I : 2.31882013e-005 J : 2.16061240e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 2645 Calibrated on : 2007-10-25 G : -1.02374143e+001 H : 1.41439052e+000 I : -1.27561930e-003 J : 1.71691149e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Fluorometer, Wetlab ECO-AFL/FL Serial number : FLRTD-297 Calibrated on : 2005-06-27 Vblank : 0.0590 Scale factor : 2.00000000e+000 7) A/D voltage 1, Free 8) A/D voltage 2, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : 758 Calibrated on : 2004-07-14 M : 21.8720 B : -1.2200 Path length : 0.250 9) A/D voltage 3, Free 10) A/D voltage 4, Altimeter Serial number : 1133 Calibrated on : Scale factor : 14.950 Offset : 0.200 11) A/D voltage 5, Oxygen, SBE 43 Serial number : 794 Calibrated on : 2007-11-06 Equation : Owens-Millard Coefficients for Owens-Millard: Soc : 3.6120e-001 Boc : 0.0000 Offset : -0.4865 Tcor : 0.0004 Pcor : 1.35e-004 Tau : 0.0 Coefficients for Murphy-Larson: Soc : 0.00000e+000 Offset : 0.00000e+000 A : 0.00000e+000 B : 0.00000e+000 C : 0.00000e+000 E : 0.00000e+000 Tau : 0.00000e+000 12) A/D voltage 6, Free 13) A/D voltage 7, Free 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : not installed Calibrated on : Conversion factor : 0.00000000 Ratio multiplier : 0.00000000 ======================================================== Plumbing configuration: Setup for VERTICAL configuration. TOP ENDCAP CONNECTIONS: JT1: (Seacable) SBE straight cable pn 17120, dn 30588-A, seacable extender. JT2: (AD0,1) SBE straight cable pn 171953, dn 32491, Chan-A: ECO-AFL, Chan-B: n/a. JT3: (AD2,3) SBE WYE cable pn 171530, dn 32528, Chan-A: WETLabs C-star, Chan-B: Biospherical PAR (Leg B for PAR has LMG-4MPD dummy plug with G-MLS-P locking sleeve installed.) JT4: (GO-1015 interface): RMG-4FSD dummy with G-FLS-P locking sleeve installed. JT5: (AD4,5) SBE WYE cable pn 172042, dn 32712, Chan-A: Benthos PSA-916, Chan-B: SBe43. JT6: (AD6,7) Impulse AG206-FSD dummy with G-FLS-P locking sleeve installed. JT7: (SBE32) SBE straight cable pn 17618, dn 305678, SBE9 to SBE32. BOTTOM ENDCAP CONNECTIONS: JB1: (Temp1) SBE straight cable pn 17086, dn 30566. JB2: (Cond1) SBE straight cable pn 17086, dn 30566. JB3: (pump) SBE WYE cable pn 17799, dn 31522a. JB4: (Temp2) SBE straight cable pn 17086, dn 30566. JB5: (Cond2) SBE straight cable pn 17086, dn 30566. JB6: (bottom contact) dummy RMG-2FSD. By Marshall Swartz. Mar 12, 2008.