EN476 Dennis McGillicuddy CTD con file Summary ================ 476a.con ===================== ctd001 ctd002 ctd003x decktest 475xapooFs1 decktest 476xapooFs1 decktest 476xapooFs2 watertest ctd777 watertest ctd778 During ctd03x - error messages at ~18.5m, 1/2Amp fuse blew, signal & ground pins shorted on Sea Cable & 9+ bulkhead connector, arcing/burning evidence. Return 9+ to SBE for repairs. swap 9+,Ts, Cs, DOs, pumps, all plumbing 476b.con ===================== ctd003-ctd020 decktest 476xapooFs3 decktest test3 watertest ctd004x watertest test watertest test2 pumps were not coming on as they should, possibly due to cond sensor swap Ts & Cs, change back to those of 476a.con 476c.con ===================== ctd021-ctd022 decktest test4 pumps were still not coming on as they should, possibly due to pump/plumbing DOs not tracking each other as well as the more freshly calibrated sensors swap DOs & pumps, C-DO-pump plumbing later - add lead to TC side of package to permanently fix flush/pump issue 476d.con ===================== ctd023-ctd114 decktest test5 decktest test6 watertest test7 watertest test8 watertest test9 watertest test10 ========================================================= ========================================================= Instrument configuration file: D:\CTDData\476a.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Primary Pump 1596 Secondary Pump 1595 Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 2899 Calibrated on : 14-Apr-10 G : 4.37878440e-003 H : 6.50381935e-004 I : 2.32263476e-005 J : 2.10143222e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 1446 Calibrated on : 13-Apr-10 G : -4.19067388e+000 H : 5.55783783e-001 I : 1.75976540e-004 J : 1.77408479e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 64853 Calibrated on : 06-Mar-09 C1 : -5.378517e+004 C2 : -3.498580e-001 C3 : 1.648580e-002 D1 : 4.036100e-002 D2 : 0.000000e+000 T1 : 2.984744e+001 T2 : -3.538190e-004 T3 : 3.972770e-006 T4 : 2.922330e-009 T5 : 0.000000e+000 Slope : 0.99984000 Offset : -1.13640 AD590M : 1.125800e-002 AD590B : -8.763490e+000 4) Frequency 3, Temperature, 2 Serial number : 2107 Calibrated on : 14-Apr-10 G : 4.14731881e-003 H : 6.27686897e-004 I : 2.08621420e-005 J : 2.11979852e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 0864 Calibrated on : 13-Apr-10 G : -4.13502391e+000 H : 5.95406684e-001 I : -6.82219193e-004 J : 6.99997404e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : 399DR Calibrated on : 25-Apr-08/11-Mar-10 field M : 20.7449 B : -1.1845 Path length : 0.250 7) A/D voltage 1, Free 8) A/D voltage 2, Altimeter Serial number : 1075 Calibrated on : 14-May-10deck Scale factor : 15.000 Offset : 0.000 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : 4479 Calibrated on : 04-May-10 M : 1.00000000 B : 0.00000000 Calibration constant : 3906300000.00000000 Multiplier : 1.00000000 Offset : -0.45813100 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 0348 Calibrated on : 10-Apr-10p Equation : Sea-Bird Soc : 4.38200e-001 Offset : -4.85100e-001 A : -1.25600e-003 B : 1.36730e-004 C : -2.94910e-006 E : 3.60000e-002 Tau20 : 1.04000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, Oxygen, SBE 43, 2 Serial number : 0343 Calibrated on : 28-Apr-10p Equation : Sea-Bird Soc : 5.69500e-001 Offset : -5.30100e-001 A : -3.05090e-003 B : 1.71330e-004 C : -3.42650e-006 E : 3.60000e-002 Tau20 : 1.45000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 12) A/D voltage 6, Free 13) A/D voltage 7, Fluorometer, Wetlab ECO-AFL/FL Serial number : 478 Calibrated on : 27-Mar-08 Vblank : 0.0210 Scale factor : 2.50000000e+001 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 20121 Calibrated on : 04-May-10 Conversion factor : 1815.88000000 Ratio multiplier : 1.00000000 ========================================================= ========================================================= Instrument configuration file: D:\CTDData\476b.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Primary Pump Secondary Pump Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 4130 Calibrated on : 13-Jan-09 G : 4.39545566e-003 H : 6.50516481e-004 I : 2.39803811e-005 J : 2.18648961e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 3220 Calibrated on : 09-Jan-09 G : -1.03780038e+001 H : 1.42645233e+000 I : 2.17267046e-004 J : 5.29591283e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 80016 Calibrated on : 06-Mar-09 C1 : -4.589988e+004 C2 : -1.257540e+000 C3 : 1.319840e-002 D1 : 3.383500e-002 D2 : 0.000000e+000 T1 : 3.008300e+001 T2 : -6.730880e-004 T3 : 3.893580e-006 T4 : 2.333810e-009 T5 : 0.000000e+000 Slope : 1.00004000 Offset : -1.73980 AD590M : 1.284520e-002 AD590B : -8.669590e+000 4) Frequency 3, Temperature, 2 Serial number : 4126 Calibrated on : 13-Jan-09 G : 4.38551263e-003 H : 6.45671874e-004 I : 2.32356573e-005 J : 2.11340376e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 2459 Calibrated on : 09-Jan-09 G : -1.02638352e+001 H : 1.51274368e+000 I : 7.31038827e-005 J : 7.50769102e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : 399DR Calibrated on : 25-Apr-08/11-Mar-10 field M : 20.7449 B : -1.1845 Path length : 0.250 7) A/D voltage 1, Free 8) A/D voltage 2, Altimeter Serial number : 1075 Calibrated on : 14-May-10deck Scale factor : 14.780 Offset : 0.000 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : 4479 Calibrated on : 08-Feb-07 M : 1.00000000 B : 0.00000000 Calibration constant : 8130100000.00000000 Multiplier : 1.00000000 Offset : -0.21722300 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 1230 Calibrated on : 16-Mar-09p Equation : Sea-Bird Soc : 4.09000e-001 Offset : -5.05100e-001 A : -2.22080e-004 B : 1.07730e-004 C : -1.79170e-006 E : 3.60000e-002 Tau20 : 1.63000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, Oxygen, SBE 43, 2 Serial number : 0345 Calibrated on : 05-Mar-09p Equation : Sea-Bird Soc : 4.27000e-001 Offset : -4.96500e-001 A : 1.94660e-004 B : 9.00850e-005 C : -1.90110e-006 E : 3.60000e-002 Tau20 : 1.08000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 12) A/D voltage 6, Free 13) A/D voltage 7, Fluorometer, Wetlab ECO-AFL/FL Serial number : 478 Calibrated on : 27-Mar-09 Vblank : 0.0210 Scale factor : 2.50000000e+001 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 20121 Calibrated on : 24-Feb-06 Conversion factor : 1700.17000000 Ratio multiplier : 1.00000000 ========================================================= ========================================================= Instrument configuration file: D:\CTDData\476c.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Primary Pump Secondary Pump Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 2899 Calibrated on : 14-Apr-10 G : 4.37878440e-003 H : 6.50381935e-004 I : 2.32263476e-005 J : 2.10143222e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 1446 Calibrated on : 13-Apr-10 G : -4.19067388e+000 H : 5.55783783e-001 I : 1.75976540e-004 J : 1.77408479e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 80016 Calibrated on : 06-Mar-09 C1 : -4.589988e+004 C2 : -1.257540e+000 C3 : 1.319840e-002 D1 : 3.383500e-002 D2 : 0.000000e+000 T1 : 3.008300e+001 T2 : -6.730880e-004 T3 : 3.893580e-006 T4 : 2.333810e-009 T5 : 0.000000e+000 Slope : 1.00004000 Offset : -1.73980 AD590M : 1.284520e-002 AD590B : -8.669590e+000 4) Frequency 3, Temperature, 2 Serial number : 2107 Calibrated on : 14-Apr-10 G : 4.14731881e-003 H : 6.27686897e-004 I : 2.08621420e-005 J : 2.11979852e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 0864 Calibrated on : 13-Apr-10 G : -4.13502391e+000 H : 5.95406684e-001 I : -6.82219193e-004 J : 6.99997404e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : 399DR Calibrated on : 25-Apr-08/11-Mar-10 field M : 20.7449 B : -1.1845 Path length : 0.250 7) A/D voltage 1, Free 8) A/D voltage 2, Altimeter Serial number : 1075 Calibrated on : 14-May-10deck Scale factor : 14.780 Offset : 0.000 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : 4479 Calibrated on : 08-Feb-07 M : 1.00000000 B : 0.00000000 Calibration constant : 8130100000.00000000 Multiplier : 1.00000000 Offset : -0.21722300 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 1230 Calibrated on : 16-Mar-09p Equation : Sea-Bird Soc : 4.09000e-001 Offset : -5.05100e-001 A : -2.22080e-004 B : 1.07730e-004 C : -1.79170e-006 E : 3.60000e-002 Tau20 : 1.63000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, Oxygen, SBE 43, 2 Serial number : 0345 Calibrated on : 05-Mar-09p Equation : Sea-Bird Soc : 4.27000e-001 Offset : -4.96500e-001 A : 1.94660e-004 B : 9.00850e-005 C : -1.90110e-006 E : 3.60000e-002 Tau20 : 1.08000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 12) A/D voltage 6, Free 13) A/D voltage 7, Fluorometer, Wetlab ECO-AFL/FL Serial number : 478 Calibrated on : 27-Mar-09 Vblank : 0.0210 Scale factor : 2.50000000e+001 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 20121 Calibrated on : 24-Feb-06 Conversion factor : 1700.17000000 Ratio multiplier : 1.00000000 ========================================================= ========================================================= Instrument configuration file: D:\CTDData\476d.con Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Primary Pump 1596 Secondary Pump 1595 Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 2899 Calibrated on : 14-Apr-10 G : 4.37878440e-003 H : 6.50381935e-004 I : 2.32263476e-005 J : 2.10143222e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 1446 Calibrated on : 13-Apr-10 G : -4.19067388e+000 H : 5.55783783e-001 I : 1.75976540e-004 J : 1.77408479e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 80016 Calibrated on : 06-Mar-09 C1 : -4.589988e+004 C2 : -1.257540e+000 C3 : 1.319840e-002 D1 : 3.383500e-002 D2 : 0.000000e+000 T1 : 3.008300e+001 T2 : -6.730880e-004 T3 : 3.893580e-006 T4 : 2.333810e-009 T5 : 0.000000e+000 Slope : 1.00004000 Offset : -1.73980 AD590M : 1.284520e-002 AD590B : -8.669590e+000 4) Frequency 3, Temperature, 2 Serial number : 2107 Calibrated on : 14-Apr-10 G : 4.14731881e-003 H : 6.27686897e-004 I : 2.08621420e-005 J : 2.11979852e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 0864 Calibrated on : 13-Apr-10 G : -4.13502391e+000 H : 5.95406684e-001 I : -6.82219193e-004 J : 6.99997404e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Transmissometer, Chelsea/Seatech/Wetlab CStar Serial number : 399DR Calibrated on : 25-Apr-08/11-Mar-10 field M : 20.7449 B : -1.1845 Path length : 0.250 7) A/D voltage 1, Free 8) A/D voltage 2, Altimeter Serial number : 1075 Calibrated on : 14-May-10deck Scale factor : 14.780 Offset : 0.000 9) A/D voltage 3, PAR/Irradiance, Biospherical/Licor Serial number : 4479 Calibrated on : 08-Feb-07 M : 1.00000000 B : 0.00000000 Calibration constant : 8130100000.00000000 Multiplier : 1.00000000 Offset : -0.21722300 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 0348 Calibrated on : 10-Apr-10p Equation : Sea-Bird Soc : 4.38200e-001 Offset : -4.85100e-001 A : -1.25600e-003 B : 1.36730e-004 C : -2.94910e-006 E : 3.60000e-002 Tau20 : 1.04000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, Oxygen, SBE 43, 2 Serial number : 0343 Calibrated on : 28-Apr-10p Equation : Sea-Bird Soc : 5.69500e-001 Offset : -5.30100e-001 A : -3.05090e-003 B : 1.71330e-004 C : -3.42650e-006 E : 3.60000e-002 Tau20 : 1.45000e+000 D1 : 1.92630e-004 D2 : -4.64800e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 12) A/D voltage 6, Free 13) A/D voltage 7, Fluorometer, Wetlab ECO-AFL/FL Serial number : 478 Calibrated on : 27-Mar-09 Vblank : 0.0210 Scale factor : 2.50000000e+001 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR/Surface Irradiance Serial number : 20121 Calibrated on : 24-Feb-06 Conversion factor : 1700.17000000 Ratio multiplier : 1.00000000 ========================================================= =========================================================