Instrument: Thiel Engineering Flowmeter Model: flowbotA S/N: D Calibration Date: 2015-10-13 Installation Date: 2016-06-30 Installation Location: Wet lab, inboard bulkhead Distance from Water Line: intake -3.7 ABL, approximately 126' aft of intake Updated: 2017-04-16 CTS RAW datafile: SSW 2016/10/03 17:01:17.092 FLOW $FLOWRATE, 70, 20566479, 146.8, 145.1, 144.9, 42.2, 1019820.000 dsLog time stamp: SSW 2016/10/03 17:01:17.092 FLOW SSW Surface Sea Water 2016/10/03 date in gmt 17:01:17.092 time in gmt FLOW data collected with flow meter data $FLOWRATE, 70, 20566479, 146.8, 145.1, 144.9, 42.2, 1019820.000 $FLOWRATE, 70, 20566479, 146.8, 145.1, 144.9, 42.2, 1019820.000 $FLOWRATE Data Tag 70 Pulse count 20566479 Cumulative Pulse Count 146.8 Current Flow Rate in ml/sec 145.1 Flow Rate for the Last Minute, ml/sec 144.9 Flow Rate for the Last Hour, ml/sec 42.2 Flow Rate for the Last 24 Hours, ml/sec 1019820.000 Elapsed Time In Seconds Since Flowmeter Was Powered Up. Notes: WMETER - Rev. 0.006 - Built at Oct 6 2016 16:49:46 BoardId: TEENSY-232 - Rev. 1.1 - S/N 1102 -------------------------------------------------- CFG settingsVersion = 1 CFG countsPerLiter = 477 --------------------------------------------------