Date: 05/14/2019 Instrument configuration file: E:\Dropbox\work\Caldata\cruise_packs\SBE09_0383_20190428_\SBE09_0383_20190514_v4.xmlcon Configuration report for SBE 911plus/917plus CTD ------------------------------------------------ Frequency channels suppressed : 0 Voltage words suppressed : 0 Computer interface : RS-232C Deck unit : SBE11plus Firmware Version >= 5.0 Scans to average : 1 NMEA position data added : Yes NMEA depth data added : No NMEA time added : No NMEA device connected to : deck unit Surface PAR voltage added : Yes Scan time added : No 1) Frequency 0, Temperature Serial number : 4230 Calibrated on : 15-Jan-19 G : 4.37272002e-003 H : 6.48950887e-004 I : 2.31795405e-005 J : 1.91673451e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 2) Frequency 1, Conductivity Serial number : 2707 Calibrated on : 24-Jan-19 G : -1.07265164e+001 H : 1.55889181e+000 I : -1.07618836e-003 J : 1.68249210e-004 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 3) Frequency 2, Pressure, Digiquartz with TC Serial number : 0383 Calibrated on : 28-Feb-19 C1 : -4.928049e+004 C2 : -5.591409e-001 C3 : 1.510530e-002 D1 : 3.944700e-002 D2 : 0.000000e+000 T1 : 3.017493e+001 T2 : -4.671701e-004 T3 : 3.967900e-006 T4 : 3.098920e-009 T5 : 0.000000e+000 Slope : 0.99997365 Offset : 3.75587 AD590M : 1.135000e-002 AD590B : -8.132450e+000 4) Frequency 3, Temperature, 2 Serial number : 4312 Calibrated on : 16-Jan-19 G : 4.38269975e-003 H : 6.51399168e-004 I : 2.32837507e-005 J : 1.87923469e-006 F0 : 1000.000 Slope : 1.00000000 Offset : 0.0000 5) Frequency 4, Conductivity, 2 Serial number : 3009 Calibrated on : 30-Jan-19 G : -9.84224434e+000 H : 1.39643112e+000 I : 3.26628364e-004 J : 4.83330059e-005 CTcor : 3.2500e-006 CPcor : -9.57000000e-008 Slope : 1.00000000 Offset : 0.00000 6) A/D voltage 0, Fluorometer, WET Labs ECO-AFL/FL Serial number : FLNTURTD-FL-1013 Calibrated on : 20180912 Dark output : 0.0900 Scale factor : 6.00000000e+000 7) A/D voltage 1, Turbidity Meter, WET Labs, ECO-NTU Serial number : FLNTURTD-TU-1013 Calibrated on : 20180912 ScaleFactor : 2.000000 Dark output : 0.084000 8) A/D voltage 2, PAR/Irradiance, Biospherical/Licor Serial number : QSP2300-70172 Calibrated on : 20130220 M : 1.00000000 B : 0.00000000 Calibration constant : 943400000.00000000 Multiplier : 1.00000000 Offset : -0.09010000 9) A/D voltage 3, Transmissometer, WET Labs C-Star Serial number : CST-1191 Calibrated on : 2019-03-06 M : 20.8460 B : -0.1250 Path length : 0.250 10) A/D voltage 4, Oxygen, SBE 43 Serial number : 0712 Calibrated on : 25-Jan-19 Equation : Sea-Bird Soc : 4.99880e-001 Offset : -5.21800e-001 A : -4.21110e-003 B : 2.02370e-004 C : -2.97100e-006 E : 3.60000e-002 Tau20 : 1.11000e+000 D1 : 1.92634e-004 D2 : -4.64803e-002 H1 : -3.30000e-002 H2 : 5.00000e+003 H3 : 1.45000e+003 11) A/D voltage 5, User Polynomial Serial number : SUNA NTR-1227 Calibrated on : 20190514 Sensor name : SUNA 2km ASY-NTR-00081 A0 : -5.43594300 A1 : 25.50267000 A2 : 0.00000000 A3 : 0.00000000 12) A/D voltage 6, Altimeter Serial number : VA500-46507 Calibrated on : 2018 Scale factor : 15.000 Offset : 0.000 13) A/D voltage 7, Fluorometer, WET Labs ECO CDOM Serial number : FLCDRTD-1963 Calibrated on : 20161118 Dark output : 0.043 Scale factor : 30.000 14) SPAR voltage, Unavailable 15) SPAR voltage, SPAR, Biospherical/Licor Serial number : QSR2200-20313 Calibrated on : 20170801 Conversion factor : 1348.30000000 Ratio multiplier : 1.00000000 Scan length : 40